RT-7600S Auto Hematology Analyzer User’s Manual
26
Fig.4-8 X-B QC list
The instrument displays 20-parameter QC result, if some parameters are not controlled, the
corresponding column shall be blank.
Upper limit of QC data columns of X-B QC list shown at a time is 200, if data are more than
200 columns, they can be viewed by switching using and ”.
Sending function can transfer some or all X-B QC data to external equipment. Operation
method is the same as data transfer in “4.1.3 L-J QC list”.
In list frame, click “Delete” button, delete confirmation dialog box pops up. After confirmed,
some or all QC data will be deleted.
4.2.3 X-B QC chart
The system provides graphic to look up X-B QC analysis data, as shown below:
Fig.4-9 X-B QC chart
Current/Total: to display current QC data point location and X-B QC analysis total.