RT-7600S Auto Hematology Analyzer User’s Manual
18
Fig.3-6 Blood sample counting screen
If “Print instantly” setting is selected, the instrument will print analysis result automatically
after test ends.
If ”Transmit instantly setting is selected, the instrument will transfer analysis result to the
computer by RS-232 serial port after test ends.
If counting and analysis operation environment is below 15 or over 35, the analysis result
of the blood sample will be unreliable; the “Temperature Low” or “Temperature high” alarm
will appear in message field of screen.
If clog aperture or air bubble appears during counting and analysis, “Aperture clogged” or “Air
bubble” alarm will appear in message field of screen.
If the remain volume of reagent is not enough, the message field will appear “Reagent
volume not enough”, it will also indicate which reagent is not enough such as diluent,
cleanser and Lyse.
If test result exceeds system output range, parameter result will be shown as “---”, then check
validity of test sample.
Note: (1) if sample splashes out of sample cup, please make necessary
protection when cleaning, never touch sample directly.
(2) Please check remaining reagent amount periodically, if
aspiration tube fails to aspirate thoroughly, please replace
reagent.
(3) Sampling needle movement may cause body injury, watch out
during aspiration.
3.4.3 Parameter prompt message
H+: indicates that parameter test result is higher than maximum of predetermined
pathogenic value.