EasyManuals Logo

REA MLV-2D User Manual

REA MLV-2D
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
REA VERIFIER
© REA Elektronik GmbH, 64367 Mühltal, Germany --- REA MLV-2D Manual Version 3 – 21/12/2015
Page 38 of 98
Device properties
This setting allows a basic choice between whether or
not to use the 90° coaxial lighting. The 90° coaxial
lighting is a paid hardware option. The 90° coaxial
lighting cannot be retrofitted but must be ordered
together with a new device. Detailed settings for the
lighting must be configured using device configuration
in TransWin32.
USB update
If an update has been downloaded (see section 16), the download also includes the
firmware for the MLV-2D. The directory "Device_Update" is contained within the
compressed, downloaded archive (ZIP file). This contains the update files (e.g.
update.rvu). Depending on the software version, variations may occur in file names
and numbers.
For the USB update, all files must be copied from this directory onto a USB pen drive.
Subdirectories must not be used. The USB pen drive does not need to be empty. Once
the function has been selected, "Application Update" and "Language Update" appear
for selection. Here, "Application Update" must always be selected. In special cases,
program text (only) can be corrected/amended by using "Language Update".

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REA MLV-2D and is the answer not in the manual?

REA MLV-2D Specifications

General IconGeneral
BrandREA
ModelMLV-2D
CategoryBarcode Reader
LanguageEnglish

Related product manuals