English 7
Apnoea
The enhanced AutoSet algorithm detects both obstructive and central apnoeas. If an apnoea occurs,
the machine responds appropriately.
Obstructive apnoea
An obstructive apnoea is when the upper airway becomes severely limited or completely obstructed.
AutoSet generally prevents obstructive apnoeas from occurring by responding to flow limitation and
snoring. If an obstructive apnoea occurs, the machine will respond by increasing pressure.
Central apnoea
During a central apnoea, the airway will remain open, but there is no flow. When a central apnoea is
detected, the machine responds appropriately by not increasing pressure.
AutoSet for Her mode
AutoSet for Her mode is based on key aspects of ResMed’s AutoSet algorithm and delivers
therapeutic responses tailored to the characteristics of female OSA patients.
The AutoSet for Her is similar to ResMed’s AutoSet algorithm with the following modifications:
Reduced rate of pressure increments designed to help prevent arousals.
Slower pressure decays.
Treats apnoeas up to 12 cm H
2
O (12 hPa) and continues to respond to flow limitation and snore
up to 20 cm H
2
O (20 hPa).
Minimum pressure (Min. Pressure) that adjusts according to the frequency of apnoeas:

Other manuals for ResMed AIRMINI

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AIRMINI and is the answer not in the manual?

ResMed AIRMINI Specifications

General IconGeneral
TypeTravel CPAP machine
Pressure Range4-20 cm H2O
Noise Level30 dBA
BluetoothYes
Operating ModeCPAP, AutoSet
Sound Level30 dBA
Altitude AdjustmentAutomatic
Operating Temperature41°F to 95°F (5°C to 35°C)
Weight0.66 lbs
Power SupplyAC 100-240V, 50-60Hz
HumidificationOptional, with HumidX/HumidX Plus waterless humidification
Data StorageVia AirMini app
Power Consumption6.3W
Ramp Time0 to 45 min (5-minute increments)
Compatible MasksAirFit P10, AirFit N20, AirFit F20
Operating Humidity10 to 95% relative humidity, non-condensing