6
Therapy information
AutoSet mode
The treatment pressure required by the patient may vary due to changes in sleep state, body
position and airway resistance. In AutoSet™ mode, the machine provides only that amount of
pressure required to maintain upper airway patency.
The machine analyses the state of the patient’s upper airway on a breath-by-breath basis and
delivers pressure within the allowed range according to the degree of obstruction. The AutoSet
algorithm adjusts treatment pressure as a function of three parameters: inspiratory flow limitation,
snore, and apnoea.
Normal airway
When the patient is breathing normally, the inspiratory flow measured by the machine as a function
of time shows a typically rounded curve for each breath.
Flow limitation
As the upper airway begins to collapse, the shape of the inspiratory flow-time curve changes. The
AirMini recognises and treats traditional as well as less common flow-limited breath wave forms.
Snore
Snoring is sound generated by vibrations of the walls of the upper airway. It is often preceded by
flow limitation or a partial obstruction of the airway.

Other manuals for ResMed AIRMINI

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AIRMINI and is the answer not in the manual?

ResMed AIRMINI Specifications

General IconGeneral
TypeTravel CPAP machine
Pressure Range4-20 cm H2O
Noise Level30 dBA
BluetoothYes
Operating ModeCPAP, AutoSet
Sound Level30 dBA
Altitude AdjustmentAutomatic
Operating Temperature41°F to 95°F (5°C to 35°C)
Weight0.66 lbs
Power SupplyAC 100-240V, 50-60Hz
HumidificationOptional, with HumidX/HumidX Plus waterless humidification
Data StorageVia AirMini app
Power Consumption6.3W
Ramp Time0 to 45 min (5-minute increments)
Compatible MasksAirFit P10, AirFit N20, AirFit F20
Operating Humidity10 to 95% relative humidity, non-condensing