English 9
Ramp Time can be set to Off, 5 to 45 minutes or Auto. When Ramp Time is set to Auto, the
machine will detect sleep onset and then gradually increase from the start pressure to the minimum
treatment pressure at a rate of 1 cm H
2
O (1 hPa) per minute. However, if sleep onset is not
detected, the machine will reach the target pressure within 30 minutes.
Expiratory Pressure Relief
Designed to make therapy more comfortable, Expiratory Pressure Relief (EPR) maintains optimal
treatment for the patient during inhalation and reduces the delivered mask pressure during
exhalation.
EPR
OnEPR is enabled.
OffEPR is disabled and not visible in Patient Mode.
The following settings are only available if EPR is On:
EPR Type
Full TimeIf set to Full Time, EPR is enabled for the whole therapy session.
Ramp OnlyIf set to Ramp Only, EPR is only enabled during ramp time.
EPR Level
1, 2, 3 cm H
2
O (1, 2, 3 hPa)
When EPR is enabled, the delivered pressure will not drop below a minimum pressure of 4 cm H
2
O
(4 hPa), regardless of the settings.
AutoSet Response
For patients who are sensitive to faster changes in pressure during therapy, AutoSet Response can
be set to either Standard or Soft. If set to soft, patients will receive gentler pressure rises during
therapy.
Patients who use the AutoSet Response feature will still get the benefits of ResMed's AutoSet
technology including improved sensitivity to flow-limitation and CSA Detection with Forced
Oscillation Technique.
SmartStart
When SmartStart™ is turned on, therapy starts automatically when the patient breathes into their
mask.
SmartStop
When SmartStop is turned on, once the patient removes their mask, therapy stops automatically
after a few seconds.
HumidX and HumidX Plus
The HumidX and HumidX Plus are waterless humidifiers. They are designed to provide more
comfort by improving the moisture level in the air the patient breathes and offering relief from
possible dryness.
The HumidX is a good starting point for finding the right level of humidification. The HumidX Plus
will improve the humidity level further. It is useful when the ambient humidity level is very low. For
example, at high altitudes or on an aircraft.
The HumidX and HumidX Plus can be used with the N20 connector for AirMini and AirFit P10 for
AirMini only.
Note: The HumidX and HumidX Plus must be replaced within 30 days of opening the
HumidX/HumidX Plus pack.

Other manuals for ResMed AIRMINI

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AIRMINI and is the answer not in the manual?

ResMed AIRMINI Specifications

General IconGeneral
TypeTravel CPAP machine
Pressure Range4-20 cm H2O
Noise Level30 dBA
BluetoothYes
Operating ModeCPAP, AutoSet
Sound Level30 dBA
Altitude AdjustmentAutomatic
Operating Temperature41°F to 95°F (5°C to 35°C)
Weight0.66 lbs
Power SupplyAC 100-240V, 50-60Hz
HumidificationOptional, with HumidX/HumidX Plus waterless humidification
Data StorageVia AirMini app
Power Consumption6.3W
Ramp Time0 to 45 min (5-minute increments)
Compatible MasksAirFit P10, AirFit N20, AirFit F20
Operating Humidity10 to 95% relative humidity, non-condensing

Related product manuals