Error Messages and Information Notes
44
9 Error Messages and Information Notes
9.1 Error Messages
Error messages inform the user about detected device or programme errors. In the event of an
error message, a fault has occurred, in which the operation of the device or the programme is
automatically interrupted. Such faults must be resolved before next startup.
Switch off the main switch and wait for
30 s before switching on again.
If the error persists, contact service.
Switch off the main switch and wait for
30 s before switching on again.
If the error persists, contact service.
Confirm the message on the control
panel.
Check the connection (control cable and
plugs) of the automatic vacuum regulation
to the AS 200 jet.
If the error persists, contact service.
Negative pressure too low
Confirm the message on the control
panel.
Check whether the industrial vacuum
cleaner is correctly connected and
generates sufficient negative pressure.
Check if the vacuum cleaner bag is full.
Check that the sieve lid is correctly
positioned on the test sieve.
If the error persists, contact service.
Confirm the message on the control
panel.
Check whether the industrial vacuum
cleaner is correctly connected and
generates sufficient negative pressure.
Check if the vacuum cleaner bag is full.
Check that the sieve lid is correctly
positioned on the test sieve.
If the error persists, contact service.
9.2 Information Notes
Notices inform the user on specific device or programme processes. The operation of the
device or programme may be interrupted briefly, but there is no fault. The information notice
must be acknowledged by the user to continue the process. Information notices provide
additional information for the user as an aid, but do not represent any device or programme
errors.
Confirm the message on the control panel.