35
7. Press the [Waveform] button to choose
CC for constant current.
8. Press the [Vector] button, once to choose
a manual vector degree. Press the [Vector]
button a second time to choose the auto
vector percentage (%) setting. Then rotate
the Central controller knob (
) to adjust to
the desired setting.
9. Press the [Beat H.] button, and then rotate
the central controller dial (
)
to set the parameter from (Beat. L) Hz to
150Hz, 1Hz/step.
10. Press the [Beat L.] button, and then
rotate the central controller dial (
) to
set the parameter from 1Hz to(Beat. H)Hz,
1Hz/step. (Note: Beat L. parameter will not
exceed Beat H. setting.
11. Press the [Time] button, and then rotate
the central controller dial (
) to set the
treatment time from 1min to 60min, 1min/step.
12. Press and hold the [Save] button until you
hear a “beep” sound if you want to save all
settings. Note: The settings will not be saved
if you have not followed the steps above.
13. Place the electrodes on the patient.
You will need to use at least one channel,
with two electrodes. You can also use two
channels with four electrodes as shown in
the figure to the left.
Place electrodes at least 2” but no more than
6” apart, per channel.
14. To adjust the output intensity, press
the corresponding channel being used for
treatment and rotate the knob clockwise
until a strong but comfortable stimulation
is felt. The “STIM” symbol indicates when
the device is giving the patient an active
stimulation. When “STIM” is not present
on the screen, DO NOT increase intensity
anymore until it is displayed again, to avoid
any sudden spikes.
15. For safety purposes, the load detection
function was designed so when the output
intensity surpasses 10.0mA/10.0V, and there
are no electrodes placed on the patient’s
skin or the device has not been connected to
electrodes, an alarm buzzer will sound and
the intensity value will flash and reset to zero.
16. To pause a channel, press the
corresponding channel knob on the top
of the device and then press the central
controller dial to pause that channel. Each
channel must be paused individually.
17. Press the [STOP] button to stop
treatment of the current channel(s). If you
want to stop the other channel(s) treatment,
press corresponding channel button first,
and then press the [STOP] button.
CAUTION: This will only stop the selected
flashing channels, not both channels.
STOP