42
5.5.4 Russian S/A Stimulation Set-up Procedure
1. To turn on the device, press the ON/OFF
switch to [
] icon which is located on the
side of the device.
2. When you turn the ComboCare™ on,
the device will perform a self-check for
about 6~8 seconds, and then the default
parameters will display the last treatment
mode.
3. Press channel button to enter channel 1
and channel 2 parameter setting mode.
4. Press the [Waveform] button to choose
the therapeutic waveform. Then rotate the
central controller dial (
) to select “Russian
S” or “Russian A” waveform.
5. Press the [Program/Save] button to
choose between P1-P10. Then rotate the
central controller dial until the desired
program number is displayed.
6. There are two modes in the device:
Normal/Default Mode and Professional Mode
To switch from Normal to Professional, press
and hold the [Step] button until you hear a
“Beep”. The program will change from P1 to
S1 for professional mode.
In Professional Mode, each program has 2
treatment steps. Press the [Step] button
again and the display will show like the figure
to the left. The parameters of each step can
be set according to the following methods.
7. Press the [Waveform] button to choose
CC for constant current.
8. Press the [Duty] button, and then rotate
the central controller dial (
) to set the
parameter from 10% to 50%, 10%/step.
Program
Save
Waveform
CC/CV
Step
Beat H.
A.M.
Duty