DG1000 Command System RIGOL
Programming Guide for DG1000
Syntax
FREQuency:STARt? [MINimum|MAXimum]
Function
Query the start frequency in sweep mode.
Return Value The query returns the start frequency set in scientific notation and
in Hz, such as: 1.000000e-06.
7.
FREQuency:STOP
Syntax
FREQuency:STOP {<frequency>|MINimum|MAXimum}
Function Set the stop frequency (used in conjunction with start frequency)
in sweep mode.
Example
FREQ:STOP MAX
8.
FREQuency:STOP?
Syntax FREQuency:STOP? [MINimum|MAXimum]
Function Query the stop frequency in sweep mode.
Return Value The query returns the stop frequency set in scientific notation and
in Hz, such as: 2.000000e+07.
9.
FREQuency:CENTer
Syntax FREQuency:CENTer {<frequency>|MINimum|MAXimum}
Function Set the center frequency (used in conjunction with frequency
span) in sweep mode.
Example
FREQ:CENT 10000000
10.
FREQuency:CENTer?
Syntax FREQuency:CENTer? [MINimum|MAXimum]
Function Query the center frequency in sweep mode.
Return Value The query returns the center frequency set in scientific notation
and in Hz, such as: 1.000000e+07.
11.
FREQuency:SPAN
Syntax FREQuency:SPAN {<frequency>|MINimum|MAXimum}
Function Set the frequency span (used in conjunction with center
frequency) in sweep mode.
Example FREQ:SPAN MAX
12.
FREQuency:SPAN?
Syntax
FREQuency:SPAN? [MINimum|MAXimum]