DG1000 Command System RIGOL
Programming Guide for DG1000
2-23
low level available and MAX selects the maximum low level
available.
Example
VOLT:LOW:CH2 MIN
16. VOLTage:LOW:CH2?
Syntax
VOLTage:LOW:CH2?
Function
Query the low level of waves output from CH2.
Return Value The query returns the low leve set in scientific notation such as:
-1.500000e+00.
17. VOLTage:OFFSet:CH2
Syntax
VOLTage:OFFSet:CH2 {<offset>|MINimum|MAXimum}
Function Set the offset voltage of CH2 in V
DC
.
Explanation <offset>is the offset voltage set by users. MIN selects the
minimum DC offset voltage for specified function and amplitude.
MAX selects the maximum value.
Example
VOLT:OFFS:CH2 MIN
18. VOLTage:OFFSet:CH2?
Syntax
VOLTage:OFFSet:CH2?
Function
Query the offset voltage of CH2.
Return Value The query returns the offset voltage set in scientific notation, such
as: -0.000000e+00.
19. VOLTage:UNIT:CH2
Syntax
VOLTage:UNIT:CH2 {VPP|VRMS|DBM}
Function
Set the unit of voltage output from CH2.
Explanation DBM could be used only in non-high resistance.
Example VOLT:UNIT:CH2 VPP
20.
VOLTage:UNIT:CH2?
Syntax
VOLTage:UNIT:CH2?
Function
Query the unit of voltage output from CH2.
Return Value The query returns VPP, VRMS or DBM.