DG1000 Command System RIGOL
Programming Guide for DG1000
2-25
Detailed information of each command:
1.
OUTPut
Syntax OUTPut {OFF|ON}
Function Disable or enable the [Output] connector of CH1 at the front panel.
The default is “OFF”.
Example
OUTP ON
2.
OUTPut?
Syntax OUTPut?
Function Query the state of the [Output] connector of CH1 at the front
panel.
Return Value The query returns OFF or ON.
3.
OUTPut:LOAD
Syntax
OUTPut:LOAD {<ohm>|INFinity|MINimum|MAXimum}
Function Select the desired output termination of CH1. The specified value
is only used for amplitude and offset voltage.
Explanations Ω is the unit of <ohm>, the default is 50Ω.
INFinitysets the output terminal asHigh Z.
Example
OUTP:LOAD 50
4.
OUTPut:LOAD?
Syntax OUTPut:LOAD? [MINimum|MAXimum]
Function Query the current load setting of CH1.
Return Value The query returns the current load setting in Ω or returns Infinity.
5.
OUTPut:POLarity
Syntax
OUTPut:POLarity {NORMal|INVerted}
Function
Set the polarity of waveform output from CH1.
Example
OUTP:POL NORM
6.
OUTPut:POLarity?
Syntax OUTPut:POLarity?
Function Query the polarity of waveform output from CH1.
Return Value The query returns NORM or INV.