EasyManua.ls Logo

Rigol DG1022

Rigol DG1022
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL DG1000 Command System
Programming Guide for DG1000
2-50
MIN=0.000001, MAX=500.
Example
BURS:INT:PER 10
6.
BURSt:INTernal:PERiod?
Syntax BURSt:INTernal:PERiod? [MINimum|MAXimum]
Function
Query the period of burst in internal trigger mode.
Return Value The query returns the burst period in scientific notation and the
default unit is s, such as: 1.000000e+01.
7.
BURSt:PHASe
Syntax BURSt:PHASe {<angle>|MINimum|MAXimum}
Function
Set the initial phase of burst.
Explanations <angle> is the phase set by users, the default unit is degree.
MIN=-180, MAX=180.
Example BURS:PHAS 150
8.
BURSt:PHASe?
Syntax
BURSt:PHASe? [MINimum|MAXimum]
Function
Query the initial phase of burst.
Return Value The query returns the initial phase of burst in scientific notation
and the default unit is degree, such as: 1.500000e+02.
9.
BURSt:STATe
Syntax
BURSt:STATe {OFF|ON}
Function
Enable or disable burst mode.
Example
BURS:STAT OFF
10.
BURSt:STATe?
Syntax BURSt:STATe?
Function Query the state of burst mode.
Return Value The query returns OFF or ON.
11.
BURSt:GATE:POLarity
Syntax
BURSt:GATE:POLarity {NORMal|INVerted}
Function Set the polarity of external gated signal at the [Ext
Trig/FSK/Burst]
conncetor on the rear panel, the default is
NORMal.

Other manuals for Rigol DG1022