Chapter 2 Command System RIGOL
DP800 Programming Guide 2-49
[:SOURce[n]]:VOLTage:PROTection[:LEVel]
Syntax [:SOURce[n]]:VOLTage:PROTection[:LEVel] {<voltage>|MINimum|MAXimum}
[:SOURce[n]]:VOLTage:PROTection[:LEVel]? [{MINimum|MAXimum}]
Description Set the overvoltage protection limit of the specified channel.
Query the overvoltage protection limit of the specified channel.
Parameter
Name Type Range Default
[n] Integer 1|2|3 1
<voltage> Real
Voltage range of the
current channel
Maximum voltage of the
current channel
Explanation When the overvoltage protection function is enabled, the output turns off
automatically when the output voltage exceeds the overvoltage protection limit
currently set. You can send the
:OUTPut:OVP:QUES? command to query
whether overvoltage protection occurs to the specified channel currently.
The function of this command is equivalent to that of the
:OUTPut:OVP:VALue
command.
When [:SOURce[n]] is omitted, the command sets the parameter of the
channel currently selected by default.
Return Format The query returns the overvoltage protection limit, for example, 8.800.
Example :VOLT:PROT 8.8
:VOLT:PROT?
Related
Commands
:OUTPut:OVP:QUES?
:OUTPut:OVP[:STATe]
:OUTPut:OVP:VALue