Chapter 3 Application Examples RIGOL
DP800 Programming Guide 3-5
To Use the Monitor
Requirement
Use the SCPI commands to realize the following functions:
Monitor CH1.
Set the monitor condition: >Volt, >Curr or >Power; set the voltage to 5V, the current to 3A and the
power to 15W; set the stop mode to OutpOff, Warning and Beeper.
Enable the monitor.
Method
1 *IDN? /*Query the ID character string of the power supply to check whether
the remote communication is normal*/
2 :INST CH1 /*Select CH1*/
3 :MONI:VOLT:COND >V,OR /*Set the voltage monitor condition to >Voltand the logic relation to
OR*/
4 :MONI:VOLT 5 /*Set the voltage of the monitor condition to 5V*/
5 :MONI:CURR:COND >C,OR /*Set the current monitor condition to “>Currand the logic relation to
OR*/
6 :MONI:CURR 3 /*Set the current of the monitor condition to 3A*/
7 :MONI:POWER:COND >P /*Set the power monitor condition to “>Power*/
8 :MONI:POWER 15 /*Set the power of the monitor condition to 15W*/
9 :MONI:STOP OUTOFF,ON /*Enable the OutpOffstop mode*/
10 :MONI:STOP WARN,ON /*Enable the Warningstop mode*/
11 :MONI:STOP BEEPER,ON /*Enable the Beeperstop mode*/
12 :MONI ON /*Enable the monitor*/

Table of Contents

Other manuals for Rigol DP800 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP800 Series and is the answer not in the manual?

Rigol DP800 Series Specifications

General IconGeneral
BrandRigol
ModelDP800 Series
CategoryPower Supply
LanguageEnglish

Related product manuals