Chapter 2 Command System RIGOL
DP800 Programming Guide 2-65
:TIMEr:ENDState
Syntax :TIMEr:ENDState {OFF|LAST}
:TIMEr:ENDState?
Description Set the end state of the timer to “Off” or “Last”.
Query the current end state of the timer.
Parameter
Name Type Range Default
{OFF|LAST} Discrete OFF|LAST OFF
Explanation The end state refers to the state of the instrument after it finishes outputting
the total number of groups of voltage/current values when the number of
cycles is a specified value.
OFF: the instrument turns off the output automatically after finishing the
output.
LAST: the instrument stops at the output state of the last group after finishing
the output.
The total number of groups in each timing output = the number of groups ×
the number of cycles; wherein, the number of groups is set by
the
:TIMEr:GROUPs command and the number of cycles is set by
the
:TIMEr:CYCLEs command.
Return Format The query returns OFF or LAST.
Example :TIME:ENDS LAST
:TIME:ENDS?
Related
Commands
:TIMEr:CYCLEs
:TIMEr:GROUPs