Processing Principles
Data store
There are two types of functions for data storing: manual store and transient
store. These are selected by setting the TYPE item in the MEMORY menu to
MANUAL or TRANSIENT.
When the store type was changed, all previous data will be cleared at the point
when the first store operation is carried out.
Manual store
Pressing the STORE key causes the currently displayed measurement results
and measurement parameters to be stored in the specified address. This ap-
plies also if the unit is currently in pause mode or if averaging is in progress.
After storing, the address number is automatically incremented by 1. After
address number 500 is reached, number 1 is selected again. If the specified
address already contains data, these will be overwritten.
Transient store
This store function is available only in analyzer mode. It stores the time
waveform continuously at a cycle that is 2.56 times the frequency span.
Store time: (256 × 500) / (frequency span × 2.56)
When the STORE key is pressed, transient store starts. After data for 500
addresses (256 data per address) have been stored, the store process is termi-
nated automatically. Storing always starts from address 1.
If the STORE key is pressed again while storing is in progress, store is termi-
nated at that point.
If the trigger function is enabled, continuous store starts when the trigger is
activated. Repeat is not performed.