Lockouts setup
Note: The Lockouts menu contains the options for liquid quality control that require the
operator to perform a liquid quality control test at specified intervals.
If the liquid quality control test is not performed correctly, or if the result is outside the
target value range, the meter is locked out from further use. The lockouts also can be set up
selectively for individual operators.
The option of setting up lockouts for operators is available only when operator lists are stored
in the meter and are activated. These lists are only available through the use of your computer
and the optional base unit.
A liquid quality control test must be completed successfully before the meter is made available
again for testing (either by the operator or in general).
Meter setup
52