Testing a capillary blood sample
64
Performing a test
8 Check the battery level. If there are
no bars left in the battery symbol, you
cannot perform any more tests.
9 Check that the date and time are
correct. Correct any wrong entries
as described in the chapter entitled
Meter setup/Setting the date.
If a lockout (OP or QC Lockout) is
displayed instead of Patient Test,
you must run a liquid quality control
before you can perform a test. (Refer
to the chapter entitled Optional Quality
control.)
When the meter is in lockout status, a
test cannot be performed.
The next steps depend on whether
you work with or without patient
lists.
Main Menu
09:15 am
09/16/2009
Control Test
Review Results
Setup
Patient Test
Main Menu
09:15 am
09/16/2009
Control Test
Review Results
Setup
QC Lockout
Logout