Error messages
96
Error Description/Solution
Error: Time for application
exceeded
The maximum allowed time of 180 seconds for applying blood
has been exceeded.
Error
Solution
Turn the meter off, remove the test strip and start a new
test using the same test strip and blood taken from a new
fingerstick from a different finger.
Error: Blood application
Error applying blood to the test strip.
Error
?
Solution
Turn the meter off and remove the test strip. Re-read the
instructions about applying blood and repeat the test with a
new test strip and blood taken from a new fingerstick from a
different finger.
Error: Measurement
Measurement error
Error
6
Solution
Turn the meter off and remove the test strip. Repeat the test
with a new test strip and blood taken from a new fingerstick
from a different finger. Do not touch or remove the test strip
when a test is in progress.
Error: Quality Control
The test strip failed the internal quality control check. The test
strip is unusable.
Error
QC
Solution
Turn the meter off and remove the test strip. Repeat the test
with a new test strip.
Troubleshooting error messages during or after blood application