EasyManua.ls Logo

Roche CoaguChek XS Plus - Cleaning;Disinfecting; Cleaning;Disinfecting the Exterior

Roche CoaguChek XS Plus
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
It is important to follow the procedures below to clean and disinfect the meter. Failure to
follow these procedures may cause malfunction of the meter.
n Do not use sprays of any sort.
n Ensure that swab or cloth is only damp, not wet.
Cleaning/Disinfecting the Exterior
n Use only the following items for cleaning/disinfecting the CoaguChek XS Plus meter
housing for a contact time of >1 minute:
- 70% isopropyl alcohol
- 10% Sodium hypochlorite solution (1 part bleach to 9 parts deionized water, made fresh
every 24 hours)
- NOTE: Do not use any other cleaning solutions/disinfectants on the
meter housing.
n Ensure that the blue test strip guide cover remains tightly closed while cleaning
the housing.
Cleaning/Disinfecting
Cleaning
87
1 Clean the Exterior
With the meter powered off, wipe the
meter’s exterior clean.
n Do not let liquid accumulate near any
opening. Make sure that no liquid
enters the meter.
2 Dry the Exterior
With a lint-free tissue, dry the meter.
n Wipe away residual moisture and
fluids after cleaning the housing.
n Allow wiped areas to dry for at least
10 minutes before performing a test.

Table of Contents

Other manuals for Roche CoaguChek XS Plus

Related product manuals