Always compare the code number you see on the display with the number that is
printed on the test strip container you are using. If the two code numbers don’t
match, insert the correct code chip in the slot in the meter.
If the code chip is missing or incorrectly inserted, error messages appear in the display.
(Please refer to the chapter Error messages.)
Testing a capillary blood sample
61
2 Always make sure that the number on
the code chip matches the number on
the label of the test strip container.
3 Slide the new code chip into the code
chip slot (as shown) until you feel it
snap into place.