If a liquid quality control test fails, an up
arrow (too high) or down arrow (too low)
is displayed and flashes.
In this display, you can scroll through
additional results or return to the main
menu.
Menu symbol:
Return to main menu
Scroll symbol:
Scroll through older values stored in
memory. (This feature is active only if
older values are present.)
When the test is complete, remove the
quality control code chip from the meter
and store it with the controls.
Quality control
82
QC Test
09:15 am
09/16/2009
0.9 INR
09/16/2009 09:15 am
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
(1.0 - 1.4 INR)
Comm.
Remove
09:15 am
09/16/2009
23 Remove the test strip from the
measurement chamber.
24 Turn the meter off.
25 Dispose of the used test strip in line with
the disposal policy of your laboratory or
medical practice.
26 Clean the meter if this becomes
necessary. (See also the chapter entitled
Cleaning.)