Loading...
Error messages
99
Error Description/Solution
Error: Memory test
(setup)
The memory contains faulty data.
Error
Solution
The meter has a fault. Please call Roche Diagnostics
Technical Service Center at 1-800-428-4674.
Error: Data transfer
Data was transferred incorrectly.
Error
Solution
Repeat the last action.
Error: Quality Control
The control solution is beyond the expiration date.
Error
EXP
Solution
Use a new control solution with the code chip belonging to it.
Error: Test memory
(capacity)
This message appears when using optional software and
memory is almost full.
Error
Solution
Connect meter to PC and download memory results.