Urisys 1100® · 5.7 · Operator's Manual · 9.0
32 7. Quality control (QC)
7. Quality control (QC)
Quality control (QC) measurements ensure the proper
functioning of the analyzer. A QC material for which the
results are known is measured and the results are then
compared against the defined ranges for these known
results.
Use commercially available urine controls, or other
suitable control material. Run a positive and negative
control at least after the weekly calibration and when a
new vial of test strips is opened.
u 6. Calibration (29)
QC measurement
The test strip handling, measurement procedure, result
reporting and data transmission in general are described
in the following section.
u 8. Reading test strips (33)
For the QC measurements follow the instruction provided
in Normal Mode (for Single Reading).
u 8.2 Performing a measurement in Normal Mode (for
Single Readings) (34)
Values obtained for these controls should fall within the
limits established by the laboratory or the manufacturer. If
the measured results obtained are outside the established
limits follow the troubleshooting actions provided in the
following section: CONTROL VALUES FALL OUTSIDE THE
DESIGNATED RANGES.
u 10. Error messages and troubleshooting (46)

Other manuals for Roche cobas Urisys 1100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas Urisys 1100 and is the answer not in the manual?

Roche cobas Urisys 1100 Specifications

General IconGeneral
BrandRoche
Modelcobas Urisys 1100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals