Roche Diagnostics
Urisys 1100® · 5.7 · Operator's Manual · 9.0
8. Reading test strips 41
8.7 Data transmission to a PC or host computer
Unidirectional The results are transmitted immediately together with the
sequence number, Patient ID (if entered), date and time.
Bidirectional The transmission can only be accomplished by using the
“Send” function when MEMORY is displayed. If a
bidirectional PC/host communication link cannot be
established, the Urisys 1100
®
system aborts transmission
after several attempts and reports an E14 INTERFACE
ERROR!.
u 10. Error messages and troubleshooting (46)
ASTM The results, sequence number, Operator ID, Device ID,
Patient ID (if entered), date and time of the measurement,
and of the last calibration will be sent to the host.
For further information on the serial interface, see:
u 11.1 Serial Interface (54)
ATTENTION:
Ensure that all required data is backed up on a
regular basis.