40 8. Reading test strips
When data are entered from the keyboard, each
character appears immediately in the display.
It can be erased by backspacing, and corrected as
necessary.
• Press the keyboard ENTER key to terminate input,
otherwise the Urisys 1100
®
system cannot start
reading.
• Press the keyboard Escape key to delete the entire
entry or turn the Urisys 1100
®
system off and then on
again.
ATTENTION:
A new Patient ID should only be entered from the
keyboard after the preceding measurement has
been completed and results printed out
(NormalMode). This allows the operator to check
correct entry directly in the display.
ATTENTION:
If a Patient ID is entered via the barcode reader
while a test strip is being read, the
Urisys 1100
®
system assigns that ID to the next
sample number in the sequence. The Patient ID
can only be erased by switching the
Urisys 1100
®
system off and then on again before
starting the next reading.
ATTENTION:
If an identification will be entered and does exceed
the maximum lengths of 13 characters for the
patient Id or the 12 characters for the operator ID
the cursor will move to the beginning (left side) of
the display and delete the previously written
characters.