Urisys 1100® · ≥5.7 · Operator's Manual · 9.0
58 12. Technical information and notices
12.2 Safety notices
This analyzer was designed and manufactured to comply
with the following international regulations “Safety
requirements for electrical equipment for measurement,
control and laboratory use” and left the factory in a safe
condition. In order to keep the analyzer in a perfect and
safe condition, it is up to the user to observe all
instructions and warnings included in this manual.
The analyzer must only be operated with the prescribed
power supply unit (Class II protection).
The analyzer is classified as Category II for overvoltage
and Degree 2 for pollution according to IEC 664.
Operating conditions: Operating In storage
Temperature: +15 ° to + 32 °C – 20 ° to + 70 °C
Relative humidity: 20 % to 80 % 20 % to 85 %
Optimum operating
conditions:
Temperature: + 20 ° to + 26 °C
Relative humidity: 30 % to 60 %
Lifetime:
Test strip tray: 18 months after first date of use
Interfaces:
PC/HOST: serial, D-Sub socket, 9-pin, female, unidirectional, bidirectional or ASTM protocol (selectable)
AT/PC keyboard: Barcode reader
5-pin DIN socket, female
Certification marks: UL, cUL
Complies with the provisions of the
applicable EU regulations.
Issued by Underwriters Laboratories, Inc.
(UL) for Canada and the US.