Roche Diagnostics
Urisys 1100® · 5.7 · Operator's Manual · 9.0
9. Cleaning and maintenance 43
9.2 Cleaning the test strip tray
When inserting and removing test strips, take care that no
urine residues come into contact with the retaining bar
mechanism.
At the end of each working day, clean the test strip tray
with water and, if necessary, with a commercial cleaning
agent and disinfectant.
r To clean the test strip tray (after each
reading)
1
To prevent carry-over and an accumulation of urinary
deposit, wipe off residues for urine from the test strip
tray with a dry, lint-free cloth after each reading.
BIOHAZARD:
Liquid waste and strip waste are potentially
biologically hazardous. Always wear gloves if
handling those materials. Dispose of the used test
strips according to the regulations for handling
potentially infectious material.