Roche Diagnostics
Urisys 1100® · ≥5.7 · Operator's Manual · 9.0
8. Reading test strips 39
8.5 Strip measurement error
If E3 STRIP MEAS. ERROR! appears, the test strip
and/or analyzer have probably been incorrectly used.
u 10. Error messages and troubleshooting (46)
8.6 Entering Patient ID, Operator ID and Authentication
Password
Patient ID When the analyzer is ready to measure (INSERT STRIP!
displayed), you may enter a Patient ID not exceeding 13
characters in length against the currently shown
sequence number by means of a barcode reader or
AT/PC keyboard.
u 11.2 Barcode reader, AT/PC keyboard (55)
The Patient ID can be verified in the display window and
entered again if necessary.
q The last Patient ID entered is stored when the START
button is pressed, i.e. when reading commences, and is
printed out and/or sent to the serial interface together
with the test result.
Operator ID If the operator ID is activated, the analyzer request the
input of the operator ID immediately after the self check
is performed when the analyzer is turned on, or when it
leaves the stand-by mode.
q You may enter an operator ID up to 12 alphanumeric
characters by means of a barcode reader or an AT/PC
keyboard.
Authentication In the authentication mode the user has to enter the
operator ID and the corresponding password in order to
have access to the instrument and the test results.
Operators with supervisor rights have access to all
results, may send the analyzer log file to the host or print
the last 10 actions of the log file and may deactivate the
authentication mode.
The list of operator IDs with corresponding passwords
may be updated from the host PC using the Download
list function key.