46 10. Error messages and troubleshooting
10. Error messages and troubleshooting
Error messages are shown in the display but are not
printed out. Following 5 minutes of inactivity, the analyzer
switches to Standby mode. The error message is
displayed again when the START button is pressed. In the
event that the analyzer is defective, contact your local
Roche Diagnostics representative.
E1 REFERENCE PAD ERROR MIDDLE!
Cause The middle portion of the reference pad on the test strip
tray is soiled or damaged.
Action 1. Switch off the analyzer.
2. Carefully clean and dry the pad.
3. Check if it is damaged (e.g. scratched, etc.).
4. Insert the test strip tray again, then wait for the self
check to finish.
5. If the error message is repeated, replace the reference
pad or use the spare test strip tray.
6. Recalibrate with Control-Test
®
M.
E15 REFERENCE PAD ERROR BOTTOM!
Cause The bottom portion of the reference pad on the test strip
tray is soiled or damaged.
Action See E1
u E1 REFERENCE PAD ERROR MIDDLE! (46)
E16 REFERENCE PAD ERROR TOP!
Cause The top portion of the reference pad on the test strip tray
is soiled or damaged.
Action See E1
u E1 REFERENCE PAD ERROR MIDDLE! (46)
E2 WRONG STRIP!
Cause The test strip used is different from the one for which the
analyzer has been programmed (Combur10Test
®
UX test
strips).