12.1 Technical data
Dimensions: Width: approx. 150 mm
Depth: approx. 290 mm
Height: approx. 95 mm
Weight: < 0,8 kg
Power supply: External mains adapter
Input: 100 - 240 V AC, 50/60 Hz, 800 mA
Output: 7.5 V DC, 3000 mA
Polarity:
Consumption: Operating: max. 15 W
Standby: 1.3 W
Noise level: 50 dB
System description: Type: reflectance photometer
Light source: 6 LEDs (light emitting diodes)
Wavelengths: 555 nm (green) 3x
610 nm (orange) 3x
Reader head: 1 head with 6 LEDs
Measuring cycle: Normal Mode: approx. 70 sec
Fast Mode: approx. 30 sec
Incubation time: 55-65 sec
Printer: thermal printer
Display: liquid crystal display, 2 lines of 24 characters
Memory: 100 samples
Date, time: integrated clock