Remote Commands
R&S
®
FSW
1040User Manual 1173.9411.02 ─ 19
Example:
DIAG:SERV:VERS?
Queries the version information.
Response:
Instrument Firmware |1.10
BIOS |FSW Analyzer BIOS V1.03-1-32-4-3 IPC10
Image Version |1.2.0
PCI-FPGA |9.01
SA-FPGA |2.43
MB-FPGA |2.0.8.0
SYNTH-FPGA |3.9.0.0
REF-FPGA |3.4.0.0
Data Sheet Version |01.00
Time Control Management |active
Analog Demod K7| |permanent
Usage:
Query only
SCPI confirmed
SYSTem:
ERRor:CLEar:ALL
This command deletes all contents of the "System Messages" table.
Example:
SYST:ERR:CLE:ALL
Usage: Event
SYSTem:ERRor:EXTended? <MessageType>[, <ChannelName>]
This command queries all system messages, or all messages of a defined type, dis-
played in the status bar for a specific measurement channel (application).
Note: This command queries the strings displayed for manual operation. For remote
programs, do not define processing steps depending on these results. Instead, query
the results of the STATus:QUEStionable:EXTended:INFO status register, which
indicates whether messages of a certain type have occurred (see "STATus:QUEStion-
able:EXTended:INFO Register" on page 637).
Query parameters:
<MessageType> ALL | INFO | WARNing | ERRor | FATal | MESSage
<ChannelName> String containing the name of the channel.
The parameter is optional. If you omit it, the command works for
the currently active channel.
Return values:
<Messages> String containing all messages of the selected type for the speci-
fied measurement channel. Each message is separated by a
comma and inserted in parentheses. If no messages are availa-
ble, empty parentheses are returned.
Example:
SYST:ERR:EXT? ALL
Returns all messages for the currently active application, e.g.
"Message 1","Message 2".