Common Measurement Settings
R&S
®
FSW
353User Manual 1173.9411.02 ─ 19
2. Connect the signal generator output to the RF INPUT connector on the front panel
of the R&S FSW.
3.
Adapt the measurement range of the R&S FSW to the filter to be tested. In this
measurement, define the following settings:
a) Press the FREQ key, select "Frequency Config" and enter "Frequency Start":
100
MHz.
b)
Enter "Frequency Stop": 300
MHz
4. Press the INPUT/OUTPUT key and select "External Generator Config".
5. In the "Interface Configuration" sub-tab, select "Generator Type":"SMU06".
6. Select "Reference: External" to synchronize the analyzer with the generator.
7. Switch to the "Measurement Configuration" sub-tab.
8. Set the "Source State" to "On".
9.
Define the generator output level as the "Source Power": -20
dBm.
10. Set the "Coupling State" to "Auto".
The "Result Frequency Start" value for the generator is indicated as 100.0
MHz.
The "Result Frequency Stop" value is indicated as 300.0
MHz.
11. Switch to the "Source Calibration" sub-tab.
12. Select the "Source Calibration Type": "Transmission" to perform a calibration
sweep and store a reference trace for the measurement setup.
Fig. 6-7: Measurement results from generator, analyzer and connecting cables
13. Select "Source Calibration Normalize": "On" to set the measurement results for the
current setup to 0, thus eliminating all effects from the generator, the analyzer and
Data Input and Output