Common Measurement Settings
R&S
®
FSW
354User Manual 1173.9411.02 ─ 19
the connecting cables from subsequent measurements with the band elimination
filter.
The reference line is displayed at 0 dB at the top of the diagram (100%).
Fig. 6-8: Normalized measurement results after calibration
Measuring the effects of the DUT
After calibration we can insert the band elimination filter (our DUT) in the measurement
setup.
1. Connect the signal generator output to the band elimination filter.
2. Connect the band elimination filter output to the RF INPUT connector on the front
panel of the R&S FSW.