Common Analysis and Display Functions
R&S
®
FSW
518User Manual 1173.9411.02 ─ 19
Peak List State............................................................................................................518
Sort Mode....................................................................................................................518
Maximum Number of Peaks........................................................................................518
Peak Excursion........................................................................................................... 518
Displaying Marker Numbers........................................................................................519
Exporting the Peak List............................................................................................... 519
Peak List State
Activates/deactivates the marker peak list. If activated, the peak list is displayed and
the peaks are indicated in the trace display.
For each listed peak the frequency/time ("X-value") and level ("Y-value") values are
given.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STAT on page 955
Sort Mode
Defines whether the peak list is sorted according to the x-values or y-values. In either
case the values are sorted in ascending order.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT on page 954
Maximum Number of Peaks
Defines the maximum number of peaks to be determined and displayed.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE on page 954
Peak Excursion
Defines the minimum level value by which a signal must rise or fall so that it will be
identified as a maximum or a minimum by the search functions.
Entries from 0 dB to 80 dB are allowed; the resolution is 0.1 dB. The default setting for
the peak excursion is 6 dB.