Remote Commands
R&S
®
FSW
699User Manual 1173.9411.02 ─ 19
[SENSe:]POWer:ACHannel:PRESet <Measurement>
This command determines the ideal span, bandwidths and detector for the current
power measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Parameters:
<Measurement> ACPower | MCACpower
ACLR measurement
CPOWer
channel power measurement
OBANdwidth | OBWidth
Occupied bandwidth measurement
CN
Carrier to noise ratio
CN0
Carrier to noise ration referenced to a 1 Hz bandwidth
Manual operation:
See "Optimized Settings (Adjust Settings)" on page 154
See "Adjust Settings" on page 187
See "Adjust Settings" on page 192
[SENSe:]POWer:ACHannel:PRESet:RLEVel
This command determines the ideal reference level for the current measurement.
This automatic routine makes sure that the that the signal power level does not over-
load the R&S FSW or limit the dynamic range by too small a S/N ratio.
To determine the best reference level, the R&S FSW aborts current measurements
and performs a series of test sweeps. After it has finished the test, it continues with the
actual measurement.
To get a valid result, you have to perform a complete sweep with synchronization to
the sweep end. This is only possible in single sweep mode.
Usage:
Event
[SENSe:]POWer:TRACe <TraceNumber>
This command selects the trace channel power measurements are performed on.
For the measurement to work, the corresponding trace has to be active.
Parameters:
<TraceNumber> Range: 1 to 6
*RST: 1
Example:
POW:TRAC 2
Assigns the measurement to trace 2.
Configuring and Performing Measurements