Parameters:
<Number> Range: 1 to 18
*RST: 1
Manual operation: See "Number of Channels (Tx, ADJ)" on page 152
11.5.3.3 Defining Weighting Filters
The following commands define weighting filters for ACLR measurements.
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel....................................................... 705
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]...............................................................705
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>................................................705
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch>................................................. 706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel......................................................706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALL...............................................................706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>.............................................. 706
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>................................................706
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel <Alpha>
This command defines the roll-off factor for the adjacent channel weighting filter.
Parameters:
<Alpha> Roll-off factor
Range: 0 to 1
*RST: 0.22
Manual operation: See "Weighting Filters" on page 158
See "Weighting Filters" on page 170
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL] <Value>
This command defines the alpha value for the weighting filter for all channels.
Parameters:
<Value> <numeric value>
*RST: 0.22
Example:
POW:ACH:FILT:ALPH:ALL 0.35
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch> <Alpha>
This command defines the roll-off factor for the alternate channel weighting filter.
Parameters:
<Alpha> Roll-off factor
Range: 0 to 1
*RST: 0.22
Manual operation: See "Weighting Filters" on page 158
See "Weighting Filters" on page 170