Remote Commands
R&S
®
FSW
714User Manual 1173.9411.02 ─ 19
If you have defined an absolute limit as well as a relative limit, the R&S FSW uses the
lower value for the limit check.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Limit> Defines the absolute limit of the specified gap channel in dBm.
<Reserved> Ignored.
Example:
CALC:LIM:ACP:GAP2:ABS 44.2dBm,0
Manual operation: See "Limit Checking" on page 172
CALCulate:
LIMit:ACPower:GAP<gap>:ABSolute:STATe <State>
This command turns the absolute limit check for the specified gap (CACLR) channel on
and off.
You have to activate the general ACLR limit check before using this command with
CALCulate:LIMit:ACPower[:STATe].
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<State> ON | OFF
*RST: OFF
Manual operation: See "Limit Checking" on page 172
CALCulate:LIMit:ACPower:GAP<gap>[:RELative] <Limit>, <Reserved>
This command defines the relative limit of the specified gap (CACLR) channel. The ref-
erence value for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S FSW uses the
lower value for the limit check.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Limit> Defines the relative limit of the specified gap channel in dB.
<Reserved> Ignored.
Example: CALC:LIM:ACP:GAP1:REL 3dB,0
Manual operation: See "Limit Checking" on page 172