Remote Commands
R&S
®
FSW
716User Manual 1173.9411.02 ─ 19
[SENSe:]POWer:ACHannel:BANDwidth|BWIDth:GAP<gap> <Bandwidth>
This command defines the bandwidth of the specified MSR gap (CACLR) channel.
This command is for MSR signals only (see CALCulate<n>:MARKer<m>:
FUNCtion:POWer:PRESet on page 700).
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Bandwidth> numeric value in Hz
*RST: 3.84 MHz
Manual operation: See "Gap (CACLR) Channel Bandwidths" on page 171
[SENSe:
]POWer:ACHannel:FILTer:ALPHa:GAP<gap> <Alpha>
This command defines the roll-off factor for the specified gap (CACLR) channel's
weighting filter.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Alpha> Roll-off factor
Range: 0 to 1
*RST: 0.22
Manual operation: See "Weighting Filters" on page 171
[SENSe:]POWer:ACHannel:FILTer:ALPHa:SBLock<sb>:CHANnel<ch> <Alpha>
This command defines the roll-off factor for the specified transmission channel's
weighting filter.
Suffix:
<sb>
.
1 | 2 | 3 | 4 | 5
sub block number
Parameters:
<Alpha> Roll-off factor
Range: 0 to 1
*RST: 0.22
Manual operation: See "Weighting Filters" on page 167
[SENSe:]POWer:ACHannel:FILTer:STATe:GAP<gap> <State>
This command turns the weighting filter for the specified gap (CACLR) channel on and
off.