Remote Commands
R&S
®
FSW
740User Manual 1173.9411.02 ─ 19
Parameters:
<Level> Absolute limit at the start frequency of a SEM range.
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STARt -10
For a detailed example see chapter 11.5.6.11, "Example: SEM
Measurement", on page 762.
Manual operation: See "Abs Limit Start/Stop <n>" on page 213
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:ABSolute:STOP
<Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STOP -15
For a detailed example see chapter 11.5.6.11, "Example: SEM
Measurement", on page 762.
Manual operation: See "Abs Limit Start/Stop <n>" on page 213
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:RELative:STARt <Limit>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit regardless of the limit check
mode.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Configuring and Performing Measurements