Remote Commands
R&S
®
FSW
739User Manual 1173.9411.02 ─ 19
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
Parameters:
<State> ON | OFF
*RST: OFF
Manual operation: See "Preamp" on page 212
[SENSe:
]ESPectrum<sb>:RANGe<range>:INSert <Mode>
This command inserts a new SEM range and updates the range numbers accordingly.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the SEM range.
Parameters:
<Mode> AFTer
Inserts a range after the selected range.
BEFore
Inserts a range before the selected range.
Manual operation: See "Insert before/after Range" on page 214
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:ABSolute:STARt
<Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
<PClass> 1..4
Power class for which the limit is defined.
Configuring and Performing Measurements