Remote Commands
R&S
®
FSW
767User Manual 1173.9411.02 ─ 19
[SENSe:]LIST:RANGe:BREak <State>
This command controls the sweep.
Parameters:
<State> ON
The R&S FSW stops after measuring one range.
To continue with the next range, you have to use INITiate:
CONMeas.
OFF
The R&S FSW sweeps all ranges in one go.
*RST: OFF
Manual operation: See "Stop After Sweep" on page 245
[SENSe:
]LIST:RANGe:COUNt?
This command queries the number of ranges in the sweep list.
Return values:
<Ranges> Number of ranges in the sweep list.
Usage: Query only
[SENSe:]LIST:RANGe<range>:DELete
This command removes a range from the sweep list.
Note that
●
you cannot delete the reference range
●
a minimum of three ranges is mandatory.
Suffix:
<range>
.
1..30
Selects the measurement range.
Usage: Event
[SENSe:]LIST:RANGe<range>:DETector <Detector>
This command selects the detector for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Configuring and Performing Measurements