Remote Commands
R&S
®
FSW
768User Manual 1173.9411.02 ─ 19
Parameters:
<Detector> NEGative
minimum peak detector
POSitive
peak detector
SAMPle
sample detector
RMS
RMS detector
AVERage
average detector
*RST: RMS
Manual operation: See "Detector" on page 244
[SENSe:]LIST:RANGe<range>:FILTer:TYPE <FilterType>
This command selects the filter type for a spurious emission measurement range.
The EMI-specific filter types are available if the EMI (R&S FSW-K54) measurement
option is installed, even if EMI measurement is not active. For details see chap-
ter 5.13.3.1, "Resolution Bandwidth and Filter Types", on page 290.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<FilterType> NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S FSW-K54) option
Return value for query is always PULS.
MIL
MIL Std (6 dB) - requires EMI (R&S FSW-K54) option
P5
5 Pole filters
*RST:
NORM
The available bandwidths of the filters are specified in the data
sheet.
Manual operation:
See "Filter Type" on page 243