[SENSe:]LIST:RANGe<range>[:FREQuency]:STARt <Frequency>
This command defines the start frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the sweep list covers, the R&S FSW will not measure the
ranges that are outside the span - results may be invalid.
greater than the span the sweep list covers, the R&S FSW will adjust the start fre-
quency of the first range and the stop frequency of the last range to the span
For more information see chapter 5.7, "Spurious Emissions Measurement",
on page 238.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Frequency> Numeric value.
*RST: -12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Manual operation: See "Range Start / Range Stop" on page 243
[SENSe:]LIST:RANGe<range>[:FREQuency]:STOP <Frequency>
This command defines the stop frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the sweep list covers, the R&S FSW will not measure the
ranges that are outside the span - results may be invalid.
greater than the span the sweep list covers, the R&S FSW will adjust the start fre-
quency of the first range and the stop frequency of the last range to the span
For more information seechapter 5.7, "Spurious Emissions Measurement",
on page 238 .
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Frequency> Numeric value.
*RST: -2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Manual operation: See "Range Start / Range Stop" on page 243