Remote Commands
R&S
®
FSW
772User Manual 1173.9411.02 ─ 19
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: 13
Default unit: dBm
Manual operation: See "Abs Limit Start/Stop" on page 246
[SENSe:]LIST:RANGe<range>:POINts[:VALue] <Points>
This command defines the number of sweep points in a spurious emission measure-
ment range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Points> For more information on sweep points see chapter 6.5.1.8, "How
Much Data is Measured: Sweep Points and Sweep Count",
on page 407.
*RST: 1001
Manual operation: See "Sweep Points" on page 245
[SENSe:]LIST:RANGe<range>:RLEVel <RefLevel>
This command defines the reference level for a spurious emission measurement
range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<RefLevel> Reference level.
Refer to the data sheet for the reference level range.
*RST: 0 dBm
Manual operation: See "Ref. Level" on page 244
[SENSe:]LIST:RANGe<range>:SWEep:TIME <SweepTime>
This command defines the sweep time for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Configuring and Performing Measurements