Remote Commands
R&S
®
FSW
781User Manual 1173.9411.02 ─ 19
Parameters:
<Time> The value range depends on the gate period you have set for
the selected trace with [SENSe:]SWEep:EGATe:TRACe:
PERiod. The following rules apply:
• the start time may not be higher than the length of the gate
• the start time may not be lower than the stop time of the gate
range of a lower order
The reset values depend on the gate range.
• for gate range 1, the start time is 0 ms
• for gate range 3, the start time is 2 ms
• for gate range 5, the start time is 4 ms
Example:
SWE:EGAT:TRAC1:STAR1 3ms
Sets the Starting point for range 1 on trace 1 at 3 ms.
Manual operation: See "Range <x> Start/Stop" on page 258
[SENSe:
]SWEep:EGATe:TRACe<k>[:STATe<range>] <State>
This command includes or excludes a gate range for a particular trace.
Suffix:
<range>
.
1...3
gate range
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SWE:EGAT:TRAC1:STAT1 ON
Activates gate range 1 for trace 1.
Manual operation: See "Gated Trigger" on page 257
See "Range <x> Use" on page 258
[SENSe:]SWEep:EGATe:TRACe<k>:STOP<range> <Time>
This command defines the stop time for a gate range.
Suffix:
<range>
.
1...3
gate range
Configuring and Performing Measurements