Remote Commands
R&S
®
FSW
780User Manual 1173.9411.02 ─ 19
11.5.8.3 Using Gate Ranges for Statistical Measurements
The following commands control gated statistical measurements.
[SENSe:]SWEep:EGATe:TRACe<k>:COMMent................................................................780
[SENSe:]SWEep:EGATe:TRACe:PERiod.........................................................................780
[SENSe:]SWEep:EGATe:TRACe<k>:STARt<range>.........................................................780
[SENSe:]SWEep:EGATe:TRACe<k>[:STATe<range>].......................................................781
[SENSe:]SWEep:EGATe:TRACe<k>:STOP<range>..........................................................781
[SENSe:]SWEep:EGATe:TRACe<k>:COMMent <Comment>
This command defines a comment for the gate of a particular trace.
Parameters:
<Comment> String containing the comment.
Example:
SWE:EGAT:TRAC1:COMM 'MyComment'
Defines a comment for the gate in trace 1.
Manual operation: See "Comment" on page 258
[SENSe:]SWEep:EGATe:TRACe:PERiod <Length>
This command defines the length of the gate for all traces.
The gate length applies to all traces.
Parameters:
<Length> Range: 100 ns to 1000 s
*RST: 2 ms
Example:
SWE:EGAT:TRAC1:PER 5ms
Defines the period for gated triggering to 5 ms.
Manual operation: See "Period" on page 258
[SENSe:]SWEep:EGATe:TRACe<k>:STARt<range> <Time>
This command defines the start time for a gate range.
Suffix:
<range>
.
1...3
gate range
Configuring and Performing Measurements