Remote Commands
R&S
®
FSW
779User Manual 1173.9411.02 ─ 19
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:STAT:CCDF ON
Switches on the CCDF measurement.
Manual operation: See "CCDF" on page 114
11.5.8.2 Configuring Statistical Measurements
The following commands configure the measurement.
Useful commands for configuring statistical measurements described elsewhere:
● [SENSe:]BANDwidth|BWIDth[:RESolution] on page 836
Remote commands exclusive to statistical measurements:
CALCulate<n>:MARKer<m>:Y:PERCent..........................................................................779
CALCulate<n>:STATistics:NSAMples.............................................................................. 779
CALCulate<n>:MARKer<m>:Y:PERCent <Probability>
This command sets a marker to a particular probability value. You can query the corre-
sponding level with CALCulate<n>:MARKer<m>:X.
Using the command turns delta markers into normal markers.
This command is available for CCDF measurements.
Parameters:
<Probability> Range: 0 % to 100 %
Example:
CALC1:MARK:Y:PERC 95PCT
Positions marker 1 to a probability of 95 %.
Manual operation: See "Percent Marker (CCDF only)" on page 256
CALCulate<n>:STATistics:NSAMples <Samples>
This command defines the number of samples included in the analysis of statistical
measurement functions.
Parameters:
<Samples> Range: Min: 100, Max: depends on the RBW filter
*RST: 100000
Example:
CALC:STAT:NSAM 500
Sets the number of measurement points to be acquired to 500.
Manual operation: See "Number of Samples" on page 257