EasyManua.ls Logo

Rohde & Schwarz FSW43 - Page 818

Rohde & Schwarz FSW43
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
818User Manual 1173.9411.02 ─ 19
Parameters:
<TimeOffset> Defines a time offset to start the measurement at the first pulse
of a trace.
*RST: 0
Default unit: s
<MeasTime> Defines the measurement time.
Default unit: s
<PulsePeriod> Defines the pulse period
<#OfPulses> Defines the number of pulses to measure.
Example:
CALC:MARK:FUNC:MSUM 50US,450US,576.9US,8
Evaluates data that contains 8 pulses during a measurement
time of 450 µs and a pulse period of 576.9 µs. The evaluation
starts with an offset of 50 µs.
[SENSe:]MPOWer:FTYPe <FilterType>
This command selects the filter type for pulse power measurements.
The EMI-specific filter types are available if the EMI (R&S FSW-K54) measurement
option is installed, even if EMI measurement is not active. For details see chap-
ter 5.13.3.1, "Resolution Bandwidth and Filter Types", on page 290.
Parameters:
<FilterType> CFILter
NORMal
P5
RRC
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S FSW-K54) option
Return value for query is always PULS.
MIL
MIL Std (6 dB) - requires EMI (R&S FSW-K54) option
[SENSe:]MPOWer:RESult[:LIST]?
This command queries the results of the pulse power measurement.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Configuring and Performing Measurements

Table of Contents

Related product manuals